Sign In | Join Free | My enlightcorp.com |
|
Brand Name : AI
Model Number : AIC-9038
Certification : ISO 9001
Place of Origin : China
MOQ : 1 set ISO 9038 Sustained Combustibility Tester
Price : Negotiable
Payment Terms : Western Union, T/T, L/C
Supply Ability : 50 sets each month
Delivery Time : 50 days
Packaging Details : ISO 9038 Sustained Combustibility Tester Packed in veneer case
Test standard : ISO 9038
Function : Determination of sustained combustibility of liquids
Test object : paints , varnishes, paint binders, solvents, petroleum or related products and adhesives
Power supply : AC 220V, 50/60Hz, 10A
Ehaust : Minimum 50L / s
Dimension : 500(W) x 300(D) x 400(H)mm
Warranty : 1 year
Weight : 20KG
Introduction:
When the hazardous liquid is heated to 0.5 ºC and 75 ºC, Sustained Combustibility of Liquids Tester is used to evaluate the continuous burning time.
Standards
ISO 9038: Determination of sustained combustibility of liquids
Features:
• Sustained Combustibility of Liquids Tester uses K- type thermocouple to measure the temperature automatically.
• Temperature controller automatic to control the temperature.
• Timer recording flame continue burning time.
• Flame can be moved to the sample position.
• Portable butane gas can be stored in the special gas tank fixture;
• Sustained Combustibility of Liquids Tester has pressure device ,which can automatically measure the current barometric pressure
• Solenoid valve to control the gas switch.
• Sustained Combustibility of Liquids Tester uses imported gas control valve control airflow to adjust the flame length.
Test results judge condition
• To be test three times. Required to record following points:
a) Check the ignition test sample or test position when the flame moves to its continuous combustion conditions
b) When the flame is placed in the test position to be tested, check whether the sample is ignited. If ignited, then the flame is returned to the "off" position, then start the time record for combustion duration of the test specimen
• If the sample is non-sustained combustion test is deemed passed, if it is sustained combustion type, so even if the failure. If any test sample that contains one of the following, it will be defined as a particular heating time duration of combustion.
• If the test sample from the test base of 45 degrees and it is ignited, the need to observe and record its combustion performance.
• When the test sample is ignited, and the test flame is moved from the test base 45 degrees position, if sustained burn more than 15 seconds, you need to observe and record the combustion performance.
Specification:
Model | AIC-9038 |
Size | 500(W) x 300(D) x 400(H)mm |
Table size | 1,000(W)×600(D)×800(H)mm |
Power | AC 220V, 50/60Hz, 10A |
Weight | 20kg |
Exhaust | Minimum 50L / s |
Auxiliary products | Butane, vacuum cleaners (optional) |